116
Using ProLink
®
Software with Micro Motion
®
Transmitters
Configure Menu: Transmitter Outputs
continued
Slug duration for RFT9739
Configuring fault
indicators for a Version 2
or earlier RFT9739
If the RFT9739 is a Version 2 or earlier model, open the Fault Indicator
list box to select from four fault indicators for milliamp outputs and the
frequency/pulse output from the RFT9739. On field-mount transmitters,
regardless of the fault indicator that is chosen, the transmitter diagnostic
LED blinks 4 times per second to indicate a fault condition.
Downscale
Milliamp outputs go to 0 mA if they produce a 0-20 mA current, or to
2 mA if they produce a 4-20 mA current; the frequency/pulse output
goes to 0 Hz.
Upscale
Milliamp outputs go to 22 mA; the frequency/pulse output goes to
15 kHz.
Last measured value
Outputs hold the values measured immediately before the fault condition
occurred.
Internal zero
If the user has scaled a milliamp output so it represents zero flow at
some setting above the minimum 0 mA or 4 mA current level, the output
goes to the scale setting that represents zero flow. The frequency/pulse
output goes to 0 Hz.
10.5 Slug duration for RFT9739
After using the Characterize dialog box to program slug flow limits (see
Section 8.5
, page 81), open the Transmitter Outputs dialog box to set a
slug duration for the RFT9739. At the Slug Duration text box, enter the
desired slug duration in seconds.
In some applications, slug flow typically occurs for short periods of time.
If the slug flow condition ceases in less than one minute, the RFT9739
can continue holding the last accurately measured flow value until
process density stabilizes within the programmed slug flow limits. The
slug duration specifies the amount of time the transmitter indicates the
last measured flow value before indicating zero flow.