Micro Motion RFT9709 Manual de Instruções

Consulte online ou descarregue Manual de Instruções para Medição, teste e controlo Micro Motion RFT9709. Micro Motion RFT9709 Instruction manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 214
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes

Resumo do Conteúdo

Página 1 - Transmitters

Using ProLink® Software with Micro Motion®TransmittersInstruction ManualNovember 1999

Página 2

viUsing ProLink® Software with Micro Motion® TransmittersContents continuedAppendixesAppendix A How to Specify the ProLink® Product . . . . . . . . .

Página 3

86Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedPressure compensation with RFT97398.7 Pressure compensa

Página 4

Using ProLink® Software with Micro Motion® Transmitters87Configure Menu: Characterize continuedPressure compensation with RFT9739Configure Menu:Transm

Página 5

88Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedPressure compensation with RFT9739Enter the new flow ca

Página 6

Using ProLink® Software with Micro Motion® Transmitters89Configure Menu: Characterize continuedPressure compensation with RFT9739Configure Menu:Transm

Página 7

90Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedPressure compensation with RFT9739Enter the new density

Página 8

Using ProLink® Software with Micro Motion® Transmitters91Configure Menu:Transmitter OutputsConfigure Menu:Transmitter InformationConfigure Menu: Event

Página 9

92Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Variables continuedOverviewFigure 9-1. Configure Transmitter Vari

Página 10 - Appendixes

Using ProLink® Software with Micro Motion® Transmitters93Configure Menu: Transmitter Variables continuedFlow unitsConfigure Menu:Transmitter OutputsCo

Página 11 - Contents

94Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Variables continuedFlow unitsTable 9-1. Mass flow measurement uni

Página 12

Using ProLink® Software with Micro Motion® Transmitters95Configure Menu: Transmitter Variables continuedFlow unitsConfigure Menu:Transmitter OutputsCo

Página 13

Using ProLink ® Software with Micro Motion® TransmittersviiContents continuedFigure 7-1 Status window for RFT9739 . . . . . . . . . . . . . . . . .

Página 14

96Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Variables continuedFlow unitsSpecial flow unitsSpecial flow units

Página 15 - 1 Before You Begin

Using ProLink® Software with Micro Motion® Transmitters97Configure Menu: Transmitter Variables continuedFlow unitsConfigure Menu:Transmitter OutputsCo

Página 16 - kit and system requirements

98Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Variables continuedFlow unitsd. Select an appropriate time unit f

Página 17 - Customer service

Using ProLink® Software with Micro Motion® Transmitters99Configure Menu: Transmitter Variables continuedFlow unitsConfigure Menu:Transmitter OutputsCo

Página 18

100Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Variables continuedDensity unitsFigure 9-6. RFT9739 for Gas6. If

Página 19 - 2 Getting Started

Using ProLink® Software with Micro Motion® Transmitters101Configure Menu: Transmitter Variables continuedDensity unitsConfigure Menu:Transmitter Outpu

Página 20 - 2.2 Communication standards

102Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Variables continuedTemperature and pressure unitsIf a density un

Página 21 - 2.3 Wiring to the transmitter

Using ProLink® Software with Micro Motion® Transmitters103Configure Menu: Transmitter Variables continuedFlow cutoffsConfigure Menu:Transmitter Output

Página 22

104Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Variables continuedFlow direction9.6 Flow directionOpen the Dire

Página 23

Using ProLink® Software with Micro Motion® Transmitters105Configure Menu: Transmitter Variables continuedInternal dampingConfigure Menu:Transmitter Ou

Página 24

viiiUsing ProLink® Software with Micro Motion® TransmittersContents continuedFigure 14-12 Temperature Slope Calibration dialog box . . . . . . 150Fig

Página 25

106Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Variables continuedInternal dampingThe damping value is the filt

Página 26 - 2.4 Connecting to the PC and

Using ProLink® Software with Micro Motion® Transmitters107Configure Menu:Transmitter OutputsConfigure Menu:Transmitter InformationConfigure Menu: Even

Página 27 - 2.5 Installing the software

108Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Outputs continuedOverviewFigure 10-1. Configure Outputs dialog b

Página 28 - Program Group dialog box

Using ProLink® Software with Micro Motion® Transmitters109Configure Menu: Transmitter Outputs continuedFrequency/pulse outputConfigure Menu:Transmitte

Página 29

110Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Outputs continuedFrequency/pulse outputTable 10-1. RFT9739 Frequ

Página 30

Using ProLink® Software with Micro Motion® Transmitters111Configure Menu: Transmitter Outputs continuedFrequency/pulse outputConfigure Menu:Transmitte

Página 31

112Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Outputs continuedMilliamp outputsK-factorThe K-factor enables th

Página 32 - 2.7 Connecting to the

Using ProLink® Software with Micro Motion® Transmitters113Configure Menu: Transmitter Outputs continuedMilliamp outputsConfigure Menu:Transmitter Outp

Página 33 - Cannot Find message

114Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Outputs continuedMilliamp outputsAdded damping on RFT9739 output

Página 34

Using ProLink® Software with Micro Motion® Transmitters115Configure Menu: Transmitter Outputs continuedFault indicators for RFT9739Configure Menu:Tran

Página 35 - hour glass

Using ProLink ® Software with Micro Motion® TransmittersixContents continuedTable 9-1 Mass flow measurement units forprocess variables . . . . . . . .

Página 36 - Communication options

116Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Outputs continuedSlug duration for RFT9739Configuring fault indi

Página 37 - Software communication

Using ProLink® Software with Micro Motion® Transmitters117Configure Menu: Transmitter Outputs continuedControl output from RFT9739Configure Menu:Trans

Página 38

118Using ProLink® Software with Micro Motion® Transmitters

Página 39 - 2.9 Exit

Using ProLink® Software with Micro Motion® Transmitters119Configure Menu:Transmitter OutputsConfigure Menu:Transmitter InformationConfigure Menu: Even

Página 40

120Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Information continuedOverviewFigure 11-1. Transmitter Informatio

Página 41 - 3 File Menu: Database

Using ProLink® Software with Micro Motion® Transmitters121Configure Menu: Transmitter Information continuedTransmitter databaseConfigure Menu:Transmit

Página 42 - Directory list box

122Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Information continuedPressure input for RFT9739Config and Calib

Página 43 - Database command buttons

Using ProLink® Software with Micro Motion® Transmitters123Configure Menu: Transmitter Information continuedPressure input for RFT9739Configure Menu:Tr

Página 44

124Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Transmitter Information continuedBurst control for RFT9739 and RFT9712/RFT97

Página 45 - 3.6 Send command

Using ProLink® Software with Micro Motion® Transmitters125Configure Menu: Transmitter Information continuedSensor database for RFT9739 and RFT9712/RFT

Página 46 - Remove command

xUsing ProLink® Software with Micro Motion® Transmitters

Página 47 - 4 File Menu: Print

126Using ProLink® Software with Micro Motion® Transmitters

Página 48 - Print setup

Using ProLink® Software with Micro Motion® Transmitters127Configure Menu:Transmitter OutputsConfigure Menu:Transmitter InformationConfigure Menu: Even

Página 49 - Transmitter connections

128Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Events continuedConfiguring event parametersFigure 12-1. Configure Events di

Página 50 - Parameter dialog box

Using ProLink® Software with Micro Motion® Transmitters129Configure Menu: Events continuedConfiguring event parametersConfigure Menu:Transmitter Outpu

Página 51 - Separator and book ends

130Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Events continuedCurrent levels for milliamp eventsEvent setpointsAny value o

Página 52 - 4.3 Print

Using ProLink® Software with Micro Motion® Transmitters131Applications Menu Help MenuConfigure Menu:Meter FactorsCalibrate Menu Test Menu13 Configure

Página 53 - 4.4 Interval print

132Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Meter Factors continuedEntering meter factorsSome applications, such as cust

Página 54 - Rate dialog box

Using ProLink® Software with Micro Motion® Transmitters133Applications Menu Help MenuConfigure Menu:Meter FactorsCalibrate Menu Test Menu14 Calibrate

Página 55 - Print file

134Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedAuto zero4. Open the Calibrate menu, then choose Auto Zero. The Flow

Página 56

Using ProLink® Software with Micro Motion® Transmitters135Calibrate Menu continuedAuto zeroApplications Menu Help MenuConfigure Menu:Meter FactorsCali

Página 57 - Error and Change Log Files

Using ProLink® Software with Micro Motion® Transmitters1File Menu: Print File Menu: Log Files View Menu: VariablesBefore You Begin Getting Started Fil

Página 58 - 5.2 Change logging

136Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedProgramming auto zero for RFT9739Diagnosing zeroing failureIf zeroin

Página 59 - Log file name

Using ProLink® Software with Micro Motion® Transmitters137Calibrate Menu continuedProgramming auto zero for RFT9739Applications Menu Help MenuConfigur

Página 60 - Change log options

138Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedDensity calibration14.4 Density calibrationDensity calibration adjus

Página 61 - 6 View Menu: Variables

Using ProLink® Software with Micro Motion® Transmitters139Calibrate Menu continuedDensity calibrationApplications Menu Help MenuConfigure Menu:Meter F

Página 62 - View Menu: Variables

140Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedDensity calibration2. Perform a two-point, one-point, or three-point

Página 63 - 6.3 Output levels window

Using ProLink® Software with Micro Motion® Transmitters141Calibrate Menu continuedDensity calibrationApplications Menu Help MenuConfigure Menu:Meter F

Página 64

142Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedDensity calibrationTable 14-3. Maximum flow rates for Micro Motion s

Página 65

Using ProLink® Software with Micro Motion® Transmitters143Calibrate Menu continuedDensity calibrationApplications Menu Help MenuConfigure Menu:Meter F

Página 66

144Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedDensity calibrationc. If possible, shut off the flow. Otherwise, pum

Página 67 - 7 View Menu: Status

Using ProLink® Software with Micro Motion® Transmitters145Calibrate Menu continuedDensity calibrationApplications Menu Help MenuConfigure Menu:Meter F

Página 68 - Table 7-1. Status indicators

2Using ProLink® Software with Micro Motion® TransmittersBefore You Begin continuedThe ProLink® kit and system requirementsUses of the ProLink® program

Página 69

146Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedDensity calibrationFigure 14-7. Density Point 1 Calibration dialog b

Página 70 - 7.3 Critical indicators

Using ProLink® Software with Micro Motion® Transmitters147Calibrate Menu continuedDensity calibrationApplications Menu Help MenuConfigure Menu:Meter F

Página 71 - Critical indicators

148Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedTemperature calibration for RFT9739Figure 14-10.Density Point 2 Cali

Página 72

Using ProLink® Software with Micro Motion® Transmitters149Calibrate Menu continuedTemperature calibration for RFT9739Applications Menu Help MenuConfig

Página 73

150Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedTemperature calibration for RFT9739Figure 14-11.Temperature Offset C

Página 74

Using ProLink® Software with Micro Motion® Transmitters151Calibrate Menu continuedMilliamp output trimApplications Menu Help MenuConfigure Menu:Meter

Página 75

152Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedMilliamp output trimFigure 14-13.Connecting a reference device to a

Página 76 - Operational indicators

Using ProLink® Software with Micro Motion® Transmitters153Calibrate Menu continuedMilliamp output trimApplications Menu Help MenuConfigure Menu:Meter

Página 77 - "Slug Flow"

154Using ProLink® Software with Micro Motion® TransmittersCalibrate Menu continuedMilliamp output trimFigure 14-16.Milliamp output trim: enter measure

Página 78 - Analog and frequency

Using ProLink® Software with Micro Motion® Transmitters155Applications Menu Help MenuConfigure Menu:Meter FactorsCalibrate Menu Test Menu15 Test Menu1

Página 79

Using ProLink® Software with Micro Motion® Transmitters3Before You Begin continuedCustomer serviceFile Menu: Print File Menu: Log Files View Menu: Var

Página 80 - 7.5 Informational indicators

156Using ProLink® Software with Micro Motion® TransmittersTest Menu continuedMilliamp output testing15.2 Milliamp output testingConnect a reference de

Página 81 - Zero indicators

Using ProLink® Software with Micro Motion® Transmitters157Test Menu continuedFrequency/pulse output testingApplications Menu Help MenuConfigure Menu:M

Página 82

158Using ProLink® Software with Micro Motion® TransmittersTest Menu continuedTest point diagnostics for Version 3 RFT9739Performing the frequency/puls

Página 83 - "Burst Mode"

Using ProLink® Software with Micro Motion® Transmitters159Test Menu continuedTest point diagnostics for Version 3 RFT9739Applications Menu Help MenuCo

Página 84 - Informational indicators

160Using ProLink® Software with Micro Motion® Transmitters

Página 85 - 8.1 Overview

Using ProLink® Software with Micro Motion® Transmitters161Applications Menu Help MenuConfigure Menu:Meter FactorsCalibrate Menu Test Menu16 Applicatio

Página 86

162Using ProLink® Software with Micro Motion® TransmittersApplications Menu continuedTotalizer controlFigure 16-1. Totalizer Control dialog box for RF

Página 87 - Flow calibration factor

Using ProLink® Software with Micro Motion® Transmitters163Applications Menu continuedApplication builderApplications Menu Help MenuConfigure Menu:Mete

Página 88

164Using ProLink® Software with Micro Motion® Transmitters

Página 89

Using ProLink® Software with Micro Motion® Transmitters165Applications Menu Help MenuConfigure Menu:Meter FactorsCalibrate Menu Test Menu17 Help Menu1

Página 90 - Field flow-calibration

4Using ProLink® Software with Micro Motion® Transmitters

Página 91

166Using ProLink® Software with Micro Motion® TransmittersHelp Menu continuedKeyboardThe ProLink Help main index consists of index headings. Each head

Página 92 - 8.3 Density factors for

Using ProLink® Software with Micro Motion® Transmitters167Help Menu continuedGetting around in HelpApplications Menu Help MenuConfigure Menu:Meter Fac

Página 93 - Density characterization

168Using ProLink® Software with Micro Motion® TransmittersHelp Menu continuedGlossary of termsJumping from one Help topic to anotherIn ProLink Help, h

Página 94 - IFT9701/IFT9703 and

Using ProLink® Software with Micro Motion® Transmitters169Appendix A How to Specify the ProLink®ProductProLink kit Product numberProLink kit with 110V

Página 95 - Slug flow limits

170Using ProLink® Software with Micro Motion® Transmitters

Página 96 - 8.6 Temperature factor for

Using ProLink® Software with Micro Motion® Transmitters171Appendix B Uploading and Downloading Configuration Files with a Model 268B.1 Uploading files

Página 97

172Using ProLink® Software with Micro Motion® TransmittersUploading and Downloading Configuration Files with a Model 268 continuedUploading filesTo sw

Página 98

Using ProLink® Software with Micro Motion® Transmitters173Uploading and Downloading Configuration Files with a Model 268 continuedUploading filesFigur

Página 99

174Using ProLink® Software with Micro Motion® TransmittersUploading and Downloading Configuration Files with a Model 268 continuedDownloading a config

Página 100 - Real-time compensation

Using ProLink® Software with Micro Motion® Transmitters175Uploading and Downloading Configuration Files with a Model 268 continuedDownloading a config

Página 101 - Compensation for stable

Using ProLink® Software with Micro Motion® Transmitters5File Menu: Print File Menu: Log Files View Menu: VariablesBefore You Begin Getting Started Fil

Página 102

176Using ProLink® Software with Micro Motion® Transmitters

Página 103

Using ProLink® Software with Micro Motion® Transmitters177Appendix C Temperature Coefficients for Flow and DensityC.1 Flow temperature coefficientThe

Página 104

178Using ProLink® Software with Micro Motion® TransmittersTemperature Coefficients for Flow and Density continuedDensity temperature coefficientTable

Página 105 - 9 Configure Menu:

Using ProLink® Software with Micro Motion® Transmitters179Temperature Coefficients for Flow and Density continuedDensity temperature coefficientTable

Página 106 - Overview

180Using ProLink® Software with Micro Motion® Transmitters

Página 107 - 9.2 Flow units

181Using ProLink® Software with Micro Motion® TransmittersAppendix D ASCII Character SetDecimal code Decimal code Decimal code000 NUL 043 + 086 V001 S

Página 108 - Flow units

182Using ProLink® Software with Micro Motion® Transmitters

Página 109

Using ProLink® Software with Micro Motion® Transmitters183Appendix E Transmitter Configuration WorksheetsThe following worksheets, shown in this appen

Página 110 - Special flow units

184Using ProLink® Software with Micro Motion® TransmittersTransmitter Configuration Worksheets continuedModel RFT9739 Configuration WorksheetDate ____

Página 111 - Transmitters

Using ProLink® Software with Micro Motion® Transmitters185Transmitter Configuration Worksheets continuedModel RFT9739 Configuration Worksheet(continue

Página 113

6Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedCommunication standards2.2 Communication standardsSwitches and jumper

Página 114 - API gravity

186Using ProLink® Software with Micro Motion® TransmittersTransmitter Configuration Worksheets continuedModel IFT9701 / IFT9703 Configuration Workshee

Página 115 - API standard volume

Using ProLink® Software with Micro Motion® Transmitters187Transmitter Configuration Worksheets continuedModel RFT9712 / RFT9729 Configuration Workshee

Página 116 - 9.5 Flow cutoffs

188Using ProLink® Software with Micro Motion® Transmitters

Página 117 - Solution:

Using ProLink® Software with Micro Motion® Transmitters189Appendix F Flowmeter Calibration RecordsThe worksheets shown in this appendix should be used

Página 118 - 9.6 Flow direction

190Using ProLink® Software with Micro Motion® TransmittersFlowmeter Calibration Records continuedModel RFT9739 Calibration RecordDate ________________

Página 119 - 9.7 Internal damping

Using ProLink® Software with Micro Motion® Transmitters191Flowmeter Calibration Records continuedModel IFT9701 / IFT9703 and RFT9712 / RFT9729 Calibra

Página 120 - Internal damping

192Using ProLink® Software with Micro Motion® Transmitters

Página 121 - Transmitter Outputs

Using ProLink ™ Software with Micro Motion® Transmitters193Page numbers in bold indicate illustrations.IndexNumerics9-volt battery 19Aaccuracy 76,

Página 122

194Using ProLink ™ Software with Micro Motion® TransmittersIndex continuedfault 53, 56, 109, 115–117, 122, 137FD 78–80field-flow calibration

Página 123 - 10.2 Frequency/pulse output

Using ProLink ™ Software with Micro Motion® Transmitters195Index continuedmA 6, 8–10, 20, 49, 56, 62–64, 69, 86, 104, 109, 112–113, 116,

Página 124 - Frequency/pulse output

Using ProLink® Software with Micro Motion® Transmitters7Getting Started continuedWiring to the transmitterFile Menu: Print File Menu: Log Files View M

Página 125 - Maximum pulse width for

196Using ProLink ™ Software with Micro Motion® TransmittersIndex continuedrack-mount 9range 2, 60, 62, 64, 112–113recalibration 149remote

Página 126 - Milliamp output variables

Using ProLink ™ Software with Micro Motion® Transmitters197Index continuedtemperature 47, 49, 64, 101–102, 109, 112–114, 128, 133calibration

Página 127 - RFT9712/RFT9729

198Using ProLink ™ Software with Micro Motion® TransmittersIndex continuedWindow menu 18Wire 60wire 58–59wiring 7–11, 20–21diagrams 7Zzero

Página 129 - 10.4 Fault indicators for

recycled paperMicro Motion Inc. USAWorldwide Headquarters7070 Winchester CircleBoulder, Colorado 80301Tel (303) 530-8400(800) 522-6277Fax (303) 530-84

Página 130 - Configuring fault

8Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedWiring to the transmitterFigure 2-2. Bell 202 temporary connection to

Página 131 - 10.6 Control output from

Using ProLink® Software with Micro Motion® Transmitters9Getting Started continuedWiring to the transmitterFile Menu: Print File Menu: Log Files View M

Página 132

10Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedWiring to the transmitterFigure 2-4. Bell 202 hard-wiring to transmi

Página 133 - Transmitter Information

Using ProLink® Software with Micro Motion® Transmitters11Getting Started continuedWiring to the transmitterFile Menu: Print File Menu: Log Files View

Página 134

12Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedConnecting to the PC and power source2.4 Connecting to the PC and po

Página 135 - 11.2 Transmitter database

Using ProLink® Software with Micro Motion® Transmitters13Getting Started continuedInstalling the softwareFile Menu: Print File Menu: Log Files View Me

Página 136 - 11.3 Pressure input for

14Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedInstalling the softwareInstallation Location dialog boxWhen the Inst

Página 137 - Pressure input for RFT9739

Using ProLink® Software with Micro Motion® Transmitters15Getting Started continuedInstalling the softwareFile Menu: Print File Menu: Log Files View Me

Página 138

Using ProLink® Software with Micro Motion®TransmittersInstruction ManualFor technical assistance, phone the Micro Motion Customer Service Department:•

Página 139

16Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedInstalling the softwareAfter ProLink software installation is comple

Página 140

Using ProLink® Software with Micro Motion® Transmitters17Getting Started continuedInstalling the softwareFile Menu: Print File Menu: Log Files View Me

Página 141 - 12 Configure Menu: Events

18Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedStart-up2.6 Start-upTo run the ProLink program, select the MMI progr

Página 142 - Configure Menu: Events

Using ProLink® Software with Micro Motion® Transmitters19Getting Started continuedConnecting to the transmitterFile Menu: Print File Menu: Log Files V

Página 143

20Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedConnecting to the transmitterTransmit light on PC Interface adaptor

Página 144 - 12.4 Reading event states

Using ProLink® Software with Micro Motion® Transmitters21Getting Started continuedConnecting to the transmitterFile Menu: Print File Menu: Log Files V

Página 145 - Meter Factors

22Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedCommunication optionsTable 2-3. Additional ProLink troubleshooting i

Página 146 - 13.3 Entering meter factors

Using ProLink® Software with Micro Motion® Transmitters23Getting Started continuedCommunication optionsFile Menu: Print File Menu: Log Files View Menu

Página 147 - 14 Calibrate Menu

24Using ProLink® Software with Micro Motion® TransmittersGetting Started continuedCommunication options6. Open the Windows Program Manager, open the M

Página 148 - Calibrate Menu

Using ProLink® Software with Micro Motion® Transmitters25Getting Started continuedExitFile Menu: Print File Menu: Log Files View Menu: VariablesBefore

Página 150 - Convergence limit

26Using ProLink® Software with Micro Motion® Transmitters

Página 151 - Zero time

Using ProLink® Software with Micro Motion® Transmitters27File Menu: Print File Menu: Log Files View Menu: VariablesBefore You Begin Getting Started Fi

Página 152 - 14.4 Density calibration

28Using ProLink® Software with Micro Motion® TransmittersFile Menu: Database continuedFile selectionFigure 3-1. Transmitter Database dialog box3.2 Fil

Página 153 - Density calibration for

Using ProLink® Software with Micro Motion® Transmitters29File Menu: Database continuedDatabase command buttonsFile Menu: Print File Menu: Log Files Vi

Página 154

30Using ProLink® Software with Micro Motion® TransmittersFile Menu: Database continuedOffline, save, and upload commands3.4 Offline, save, and upload

Página 155 - Table 14-2. Density of air

Using ProLink® Software with Micro Motion® Transmitters31File Menu: Database continuedLoad commandFile Menu: Print File Menu: Log Files View Menu: Var

Página 156

32Using ProLink® Software with Micro Motion® TransmittersFile Menu: Database continuedRemove commandTo send a configuration file to the transmitter:1.

Página 157 - Table 14-4. Density of water

Using ProLink® Software with Micro Motion® Transmitters33File Menu: Print File Menu: Log Files View Menu: VariablesBefore You Begin Getting Started Fi

Página 158 - IFT9701/IFT9703

34Using ProLink® Software with Micro Motion® TransmittersFile Menu: Print continuedPrint setupFigure 4-1. Print Setup/Ticket Builder dialog boxSelect,

Página 159

Using ProLink® Software with Micro Motion® Transmitters35File Menu: Print continuedPrint setupFile Menu: Print File Menu: Log Files View Menu: Variabl

Página 160

Using ProLink ® Software with Micro Motion® TransmittersiContents1 Before You Begin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.

Página 161

36Using ProLink® Software with Micro Motion® TransmittersFile Menu: Print continuedPrint setup2. At the Add Transmitter Tag dialog box, use the Tag Na

Página 162 - 14.5 Temperature calibration

Using ProLink® Software with Micro Motion® Transmitters37File Menu: Print continuedPrint setupFile Menu: Print File Menu: Log Files View Menu: Variabl

Página 163

38Using ProLink® Software with Micro Motion® TransmittersFile Menu: Print continuedPrintClick an option button to choose a separator. If Other is chos

Página 164

Using ProLink® Software with Micro Motion® Transmitters39File Menu: Print continuedInterval printFile Menu: Print File Menu: Log Files View Menu: Vari

Página 165 - Preparing for milliamp

40Using ProLink® Software with Micro Motion® TransmittersFile Menu: Print continuedUpdate rate4.5 Update rateUpdate Rate controls how often the ProLin

Página 166 - Performing milliamp

Using ProLink® Software with Micro Motion® Transmitters41File Menu: Print continuedPrint fileFile Menu: Print File Menu: Log Files View Menu: Variable

Página 167

42Using ProLink® Software with Micro Motion® Transmitters

Página 168

Using ProLink® Software with Micro Motion® Transmitters43File Menu: Print File Menu: Log Files View Menu: VariablesBefore You Begin Getting Started Fi

Página 169 - 15 Test Menu

44Using ProLink® Software with Micro Motion® TransmittersFile Menu: Error and Change Log Files continuedChange loggingError log optionsUse the Error L

Página 170 - Milliamp output testing

Using ProLink® Software with Micro Motion® Transmitters45File Menu: Error and Change Log Files continuedChange loggingFile Menu: Print File Menu: Log

Página 171 - Calibrate Menu Test Menu

iiUsing ProLink® Software with Micro Motion® TransmittersContents continued4 File Menu: Print. . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 172

46Using ProLink® Software with Micro Motion® TransmittersChange log optionsUse the Change Log check boxes to set the options for the change log file.

Página 173 - Test Menu

Using ProLink® Software with Micro Motion® Transmitters47File Menu: Print File Menu: Log Files View Menu: VariablesBefore You Begin Getting Started Fi

Página 174

48Using ProLink® Software with Micro Motion® TransmittersView Menu: Variables continuedProcess variables windowFigure 6-1. Process variables window fo

Página 175 - 16 Applications Menu

Using ProLink® Software with Micro Motion® Transmitters49View Menu: Variables continuedOutput levels windowFile Menu: Print File Menu: Log Files View

Página 176 - Applications Menu

50Using ProLink® Software with Micro Motion® TransmittersView Menu: Variables continuedCopying displayed values to other software applicationsFigure 6

Página 177 - 16.3 Application builder

Using ProLink® Software with Micro Motion® Transmitters51View Menu: Variables continuedCopying displayed values to other software applicationsFile Men

Página 178

52Using ProLink® Software with Micro Motion® Transmitters

Página 179 - 17 Help Menu

Using ProLink® Software with Micro Motion® Transmitters53Configure Menu:Transmitter OutputsConfigure Menu:Transmitter InformationConfigure Menu: Event

Página 180 - 17.5 Context-sensitive Help

54Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedOverviewTable 7-1. Status indicatorsFigure 7-1. Status window for

Página 181 - Getting around in Help

Using ProLink® Software with Micro Motion® Transmitters55View Menu: Status continuedOverviewConfigure Menu:Transmitter OutputsConfigure Menu:Transmitt

Página 182 - 17.7 Glossary of terms

Using ProLink ® Software with Micro Motion® TransmittersiiiContents continued7.5 Informational indicators . . . . . . . . . . . . . . . . . . . . . .

Página 183

56Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedFault outputs7.2 Fault outputsIf a flowmeter failure occurs, trans

Página 184

Using ProLink® Software with Micro Motion® Transmitters57View Menu: Status continuedCritical indicatorsConfigure Menu:Transmitter OutputsConfigure Men

Página 185 - B.1 Uploading files

58Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedCritical indicatorsTo troubleshoot these status indicators, follow

Página 186 - Uploading files

Using ProLink® Software with Micro Motion® Transmitters59View Menu: Status continuedCritical indicatorsConfigure Menu:Transmitter OutputsConfigure Men

Página 187

60Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedCritical indicatorsTable 7-4. Nominal resistance and voltage range

Página 188 - B.2 Downloading a

Using ProLink® Software with Micro Motion® Transmitters61View Menu: Status continuedCritical indicatorsConfigure Menu:Transmitter OutputsConfigure Men

Página 189

62Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedOperational indicators"Analog Input Error" and "Pre

Página 190

Using ProLink® Software with Micro Motion® Transmitters63View Menu: Status continuedOperational indicatorsConfigure Menu:Transmitter OutputsConfigure

Página 191 - Flow and Density

64Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedOperational indicatorsTo program slug flow limits, follow these st

Página 192

Using ProLink® Software with Micro Motion® Transmitters65View Menu: Status continuedOperational indicatorsConfigure Menu:Transmitter OutputsConfigure

Página 193

ivUsing ProLink® Software with Micro Motion® TransmittersContents continued10Configure Menu: Transmitter Outputs. . . . . . . 10710.1 Overview . . .

Página 194

66Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedInformational indicators"Raw Flow Overflow" and "Ra

Página 195

Using ProLink® Software with Micro Motion® Transmitters67View Menu: Status continuedInformational indicatorsConfigure Menu:Transmitter OutputsConfigur

Página 196

68Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedInformational indicatorsTable 7-8. Troubleshooting informational f

Página 197 - Worksheets

Using ProLink® Software with Micro Motion® Transmitters69View Menu: Status continuedInformational indicatorsConfigure Menu:Transmitter OutputsConfigur

Página 198

70Using ProLink® Software with Micro Motion® TransmittersView Menu: Status continuedInformational indicatorsEvent indicatorsWith flow, density, temper

Página 199

Using ProLink® Software with Micro Motion® Transmitters71Configure Menu:Transmitter OutputsConfigure Menu:Transmitter InformationConfigure Menu: Event

Página 200

72Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedOverviewTo open the Characterize Sensor dialog box, ope

Página 201

Using ProLink® Software with Micro Motion® Transmitters73Configure Menu: Characterize continuedFlow calibration factorConfigure Menu:Transmitter Outpu

Página 202

74Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedFlow calibration factorTable 8-1. Temperature coefficie

Página 203 - • RFT9379 Calibration Record

Using ProLink® Software with Micro Motion® Transmitters75Configure Menu: Characterize continuedFlow calibration factorConfigure Menu:Transmitter Outpu

Página 204 - Flowmeter Calibration Records

Using ProLink ® Software with Micro Motion® TransmittersvContents continued14Calibrate Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 205

76Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedFlow calibration factorSensor or transmitter replaced i

Página 206

Using ProLink® Software with Micro Motion® Transmitters77Configure Menu: Characterize continuedFlow calibration factorConfigure Menu:Transmitter Outpu

Página 207

78Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedDensity factors for RFT973910.Multiply the meter factor

Página 208

Using ProLink® Software with Micro Motion® Transmitters79Configure Menu: Characterize continuedDensity factors for RFT9739Configure Menu:Transmitter O

Página 209

80Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedDensity factor for IFT9701/IFT9703 and RFT9712/RFT9729T

Página 210

Using ProLink® Software with Micro Motion® Transmitters81Configure Menu: Characterize continuedSlug flow limitsConfigure Menu:Transmitter OutputsConfi

Página 211

82Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedTemperature factor for RFT97398.6 Temperature factor fo

Página 212

Using ProLink® Software with Micro Motion® Transmitters83Configure Menu: Characterize continuedTemperature factor for RFT9739Configure Menu:Transmitte

Página 213

84Using ProLink® Software with Micro Motion® TransmittersConfigure Menu: Characterize continuedTemperature factor for RFT97398. Using the values deter

Página 214 - Micro Motion Asia

Using ProLink® Software with Micro Motion® Transmitters85Configure Menu: Characterize continuedTemperature factor for RFT9739Configure Menu:Transmitte

Comentários a estes Manuais

Sem comentários