10
RFT9709 Transmitter Instruction Manual
Wiring
continued
Cable types
Micro Motion supplies 9-wire jacketed, shielded, or armored cable, as
illustrated in Figure 2-3.
• All cable types are acceptable for cable tray installation.
• All cable types must have the jacket extend under the wiring terminals
access cover, as illustrated in Figure 2-4.
• Shielded and armored cable must have the braided shield terminated
as illustrated in Figure 2-4.
Figure 2-3.
Cable types
Figure 2-4.
Terminating flowmeter
cable
Braided shield
Jacketed cable
Shielded or
armored cable
Jacket
Outer jacket
Jacket
For shielded or
armored flowmeter
cable, connect
braided shield at
mounting boss
Jacket must extend
under access cover
Mounting boss
Braided shield
09wire.fm Page 10 Monday, March 1, 1999 4:31 PM